Leading provider of Thermal Conductivity Testing and Thermal
Conductivity Instruments as well as Thermal Resistivity,
Specific Heat Capacity and Thermal Diffusivity.
The Hot Disk Transient Plane Source TPS 2500 S constitutes an industry benchmark Thermal Conductivity test system. The TPS 2500 S Thermal Conductivity Instrument is designed for analyzing thermal transport properties of solids, liquids, paste and powders, including various types of geometry and dimensions. The TPS 2500 S and TPS 1500 Thermal Conductivity Systems are the only Transient Plane Source (TPS) Method to meet a testing standard (ISO/DIS 22007-2.2).
The patented Hot Disk TPS technique, which is celebrating over twenty years of development and more than 200 clients worldwide– is highly appreciated for its ability to simultaneously determine thermal conductivity, thermal diffusivity and specific heat capacity from a single measurement, with limited emphasis on sample preparation. The operative simplicity of the TPS 2500 S and the performance achieved when using optimally designed TPS sensors, makes this system a robust and ideal tool for tests of solids, liquids, powders, pastes and foams. The trademark high accuracy and precision makes the instrument suitable for QC testing.
Optional software modules allows the instrument to be used in special applications, such as measurements
of thin films, coatings or adhesive layers (Thin Films module), high-conducting sheets or slabs (Slabs module), and anisotropic samples or layered structures (Anisotropic module) – important applications in the
electronics, geo-thermal, building materials, nuclear and chemical industries.
The test & analysis software for the Hot Disk TPS 2500 S system incorporates tools for automated measurements as well as automatic temperature control of external devices. The software also incorporates tools for exporting results to 3rd-party software (MS Excel), for additional processing or statistical analysis.
TPS 2500 S THERMAL CONDUCTIVITY METER SPECIFICATIONS
| Materials | Solids, Liquids, Powders & Paste |
| Thermal Conductivity | 0.005 to 500 W/mK |
| Thermal Diffusivity | 0.1 to 100 mm2/s |
| Specific Heat Capacity | Up to 5 MJ/m3K |
| Measurement Time | 1 to 1280 seconds |
| Reproducibility | Typically better than 1% |
| Accuracy | Better than 5% |
| Temperature Range | Cryogenic to 750°C |
| Power Requirements | Adjusted to the line voltage in the country of use |
| Smallest Sample Dimensions | 0.01mm High, 2mm Diameter (Cylindrical sample) |
| Largest Sample Size | Unlimited |
| Sensor Types Available | Kapton insulated with or without cable (from cryogenic temperatures up to 180°C). |
| Mica insulated without cable (Room Temp. up to 750°C). | |
| Meets ISO Standard | ISO/DIS 22007-2.2 |