Method Comparison


Thermal Conductivity Method Comparison
Transient Plane Source TPS 2500 S & TPS 1500Modified Transient Plane SourceLaser FlashHeat-Flow Meter
ASTM/ISO/DIN StandardizationYesNoYesYes
Direct Thermal Conductivity MeasurementYesNoNoYes
Can Measure Isotropic & Non-Isotropic MaterialsYesNoNoNo
Requires - Interface Liquid Between Sensor & SampleNoYesNoYes
Requires - OEM CalibrationNoYesNoYes
Test Time0.005 - 1280 sec0.8 - 5 secN/AN/A
Speed & Flexibility
Sample PreparationNoneNoneModerateExtensive
Time ResultsSecondsSecondsMinutesHours
Training TimeOne-dayOne-dayOne-dayOne-day
Non-DestructiveYesNo (liquid interface)No ( interface coating)No (liquid interface)
Easy to Export ResultsYesYesYesYes
Ranges
Thermal Conductivity0.005 - 500 w/mk0.1 - 100 w/mk0.1 - 200 w/mk0.01 - 2 w/mk
Thermal Diffusivity0.1 - 100 mm2/sNot capable0.01 - 100 mm2/sNot capable
Specific HeatUp to 5 MJ/m3kNot capableNot capableNot capable
EffusivityYesYesNoNo
Temperature-100°C - 1000°C-20°C - 200°C-100°C - 500°C0°C - 200°C
Penetration Thickness0.01 - 30 mm1-5 mm0.05 - 5 mm100 mm
Smallest Size2 mm diameter or square17 mm diameter8 mm square 300 mm square
Dual + Single Sided TestingYesNoNoNo
Materials for Thermal ConductivitySolids, Liquids, Pastes PowdersSolids, Liquids, PastesSolids, LiquidsSolids